Preparation of aqueous standards for low temperature x-ray microanalysis

A technique, using Nuclepore polycarbonate membrane filters as a containing medium for very small volumes of ionic standard solutions, to produce homogeneous ice standards is described. The standards are suitable for use in a scanning electron microscope. The relationship between elemental X-ray counts and ionic concentration is found to be linear. The method is rapid and simple. Minimum detectable concentrations are given.

Details

Publication status:
Published
Author(s):
Authors: Reid, A.P., Potts, W.T.W., Oates, K., Mulvaney, R. ORCIDORCID record for R. Mulvaney, Wolff, E.W.

On this site: Eric Wolff, Robert Mulvaney
Date:
1 January, 1992
Journal/Source:
Microscopy Research and Technique / 22
Page(s):
207-211
Link to published article:
https://doi.org/10.1002/jemt.1070220210